• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Covidien Sonicision; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Covidien Sonicision; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number SCDA39
Medical Device Problem Codes Failure to Deliver Energy (1211); Peeled/Delaminated (1454); Failure to Cut (2587); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during procedure, the device presented a defect in cut and coagulation and the teflon presented failure detaching from the jaws of the device.The device was immediately replaced.There was no patient harm/injury.Photo observation showed the jaw liner sliding midway of the jaw.Medtronic's initial evaluation of the incident device found the jaw liner was removed.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functional testing found that the jaw liner damage was caused by the device jaws being clamped and activated multiple times with too little or no tissue present in the jaws.Extended activation under this condition will cause the jaw liner to melt and become damaged.In this case the jaw liner was likely removed while handling/cleaning the device prior to shipping it to the returned product analysis lab.It was reported that the device presented a defect in cut and coagulation and the teflon presented failure detaching from the jaws of the device.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: not to activate the instrument with the clamping jaw closed unless there is tissue present, as doing so may damage the device jaws.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sonicision
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
Covidien
6044 longbow court
boulder CO 80301 -329
Manufacturer (Section G)
Covidien
6044 longbow court
boulder CO 80301 -329
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112--439
+1(800)646-4633
MDR Report Key50903722
Report Number1717344-2026-01301
Device Sequence Number6001804
Product Code LFL
UDI-Device Identifier10884521592841
UDI-Public(01)10884521592841(17)301028(10)53020239X
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K180149
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSCDA39
Device Catalogue NumberSCDA39
Device Lot Number53020239X
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/25/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-