| Model Number |
SCDA39 |
| Medical Device Problem Codes |
Failure to Deliver Energy (1211); Peeled/Delaminated (1454); Failure to Cut (2587); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during procedure, the device presented a defect in cut and coagulation and the teflon presented failure detaching from the jaws of the device.The device was immediately replaced.There was no patient harm/injury.Photo observation showed the jaw liner sliding midway of the jaw.Medtronic's initial evaluation of the incident device found the jaw liner was removed.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functional testing found that the jaw liner damage was caused by the device jaws being clamped and activated multiple times with too little or no tissue present in the jaws.Extended activation under this condition will cause the jaw liner to melt and become damaged.In this case the jaw liner was likely removed while handling/cleaning the device prior to shipping it to the returned product analysis lab.It was reported that the device presented a defect in cut and coagulation and the teflon presented failure detaching from the jaws of the device.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: not to activate the instrument with the clamping jaw closed unless there is tissue present, as doing so may damage the device jaws.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|