Medline Industries, LP MEDLINE; ANEROID,HANDHELD,PREMIER,ADULT Back to Search Results
Catalog Number MDS9410
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported receiving a medline handheld aneroid that arrived with a hole in the tubing.Medline received the sample in like new condition, performed a visual inspection, and attempted to test the sample by inflating to 300mmhg and observing for any other signs of defect/leaking; the cuff was able to inflate and hold a pressure of 300mmhg without any significant dropping in pressure.The complaint was confirmed, and the reported hole in the tubing near the gauge was the only defect.The cause was documented as cannot determine cause with received sample, and the division will continue to monitor the issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Event or Problem Description
It was reported that "this arrived with a hole in the tubing.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
ANEROID,HANDHELD,PREMIER,ADULT
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
sasidhar lakku
three lakes drive
northfield, IL 60093--275
MDR Report Key50904163
Report Number1417592-2026-51526
Device Sequence Number6001851
Product Code DXQ
UDI-Device Identifier00080196773193
UDI-Public00080196773193
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K211084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS9410
Device Lot Number65124080002
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/29/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2024
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-