| Model Number |
DDPB3D4 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Correction: b5.Desc evt problem product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during implantation, there were multiple attempts to insert the right atrial (ra) lead into the header however, noise was observed and persisted on the atrial channel.It was further reported that there were intermittent connector/header issues and the physician requested a new implantable cardioverter defibrillator (icd).The atrial lead noise was not an issue with the new device.The ra lead was implanted, and the icd was not implanted.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|