"introduced a 24fr gore dry seal on the patients right.Used ivus to determine proper landing zone in the arch.36x250 relay pro was prepped with saline verified control knob was in position 1 and introduced in patient.Device was then removed to perform angiogram with contrast.Device was introduced to distal landing zone and dr chou started deploying graft by turning deployment grip away from him.The inner sheath advanced so that a little more than half of the stent was out of the outer sheath.At this point device would not advance anymore.Dr chou pressed the disengagement button and tried to advance inner sheath, and there was no movement.At this point we noticed that the control knob was slightly out of the pocket in position one.Dr chou adjusted that and we still had no luck in advancing device to the proper landing zone in the arch.At this point the device was put in position 2 and removed from patient.Dr chou then proceeded to put 2 gore thoracic devices in and finish the case.The surgical tech had the relay pro on the back table and in going from position 2 and then position 1 and back to position 2 while turning deployment grip, the inner sheath finally was out of the outer sheath.We then tried to remove the inner sheath with various changes to the control knob and movements, and we could not get the inner sheath back to deploy the graft completely."
patient outcome: "patient was fine after deployment of the gore devices.".
|
|
Bolton medical, inc.(d/b/a terumo aortic), herein known as the "company", is submitting this report pursuant to 21 cfr part 803, has made reasonable efforts to obtain complete information, and has provided as much as is available to the company as of the submission date of this report.This report is based on information obtained by the company, which may not have been able to fully investigate or verify prior to the date the report was required by the regulations.Any required fields that are unpopulated are blank because the information is currently unknown, unavailable, or not applicable.This report does not constitute an admission or a conclusion by anyone that the device caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the certain regulations, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|