Block b3: exact date unknown, event occurred two years ago from the date the manufacturer was made aware.Block d.2b; additional applicable product code: lgw, qrb
additional suspect medical device component involved in the event:
model number/catalog number: sc-2317-70,
serial number: (b)(6),
batch/lot number: 7083020,
model/catalog description: infinion cx lead kit, 70cm,
unique identifier (udi) #: (b)(4).Sc-2317-70 (sn (b)(6))
investigation results, media review, device analysis: the device was not returned for analysis; therefore, a technical analysis could not be performed.The photo provided by the customer confirms that there is high impedance.Device history record: it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.Sc-2317-70 (sn (b)(6))
investigation results, media review, device analysis: the device was not returned for analysis; therefore, a technical analysis could not be performed.According to media inspection of the picture provided by the customer, high impedances were verified.The complaint was confirmed.Device history record: it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|