Model Number SC-2317-70
Medical Device Problem Codes High impedance (1291); Migration (4003)
Health Effect - Clinical Code Fall (1848)
Date of Event 03/07/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block b3: exact date unknown, event occurred two years ago from the date the manufacturer was made aware.Block d.2b; additional applicable product code: lgw, qrb additional suspect medical device component involved in the event: model number/catalog number: sc-2317-70, serial number: (b)(6), batch/lot number: 7083020, model/catalog description: infinion cx lead kit, 70cm, unique identifier (udi) #: (b)(4).Sc-2317-70 (sn (b)(6)) investigation results, media review, device analysis: the device was not returned for analysis; therefore, a technical analysis could not be performed.The photo provided by the customer confirms that there is high impedance.Device history record: it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.Sc-2317-70 (sn (b)(6)) investigation results, media review, device analysis: the device was not returned for analysis; therefore, a technical analysis could not be performed.According to media inspection of the picture provided by the customer, high impedances were verified.The complaint was confirmed.Device history record: it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Event or Problem Description
It was reported that after a fall, it was found out that the right lead had two impedances and the lead had possibly migrated but was unable to confirm it.After program optimization, the patient got good coverage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Infinion? CX
Common Device Name
stimulator, spinal-cord, totally implanted for pain relief
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
Boston Scientific Corporation
road 698, lot no. 12
dorado PR 00646 -260
PR   00646-2602
Manufacturer Contact
ricardo amaro castro
25155 rye canyon loop
valencia, CA 91355
+1(320)800-7570
MDR Report Key50910714
Report Number3006630150-2026-07137
Device Sequence Number5330277
Product Code LGW
UDI-Device Identifier08714729861638
UDI-Public(01)08714729861638(17)250801(10)7083112
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/01/2025
Device Model NumberSC-2317-70
Device Catalogue NumberSC-2317-70
Device Lot Number7083112
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/28/2026
Date Device Manufactured08/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age83 YR
Patient SexMale
-
-