Other text: the returned product was evaluated.The device powered on using both battery and ac power.However, the display became inactive, followed by an alarm condition consisting of one long beep and two short beeps, while the home button flashed red continuously.An internal inspection identified a defective solder joint at the u15 chip on the mx-7 tech board, which prevented completion of the power-on sequence.Additionally, the device log files also recorded an ¿mmxbi board failure" error message.The reported issue was confirmed.Health effect impact code: no adverse event; no patient impact.
|