| Model Number |
430100-02 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during central processing, the user reported that the tip of a single port monopolar curved scissors instrument was broken.There was no report of patient involvement or patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the single port monopolar curved scissors instrument to perform failure analysis (fa).Fa was able to confirm/reproduce the reported the complaint.The instrument was found to have a broken instrument tube adapter.This component is located at the distal end of the instrument and serves as the interface between the instrument and the user-installed tip accessory.The broken piece was not returned and measures approximately 1.78 mm x 3.03 mm in size.The housing was removed and found no damage.The inputs were manually articulated and passed.The probable root cause of a damaged tube adapter is attributed to either tip accessory installation issues or damage during use.
|
| |
|