| Model Number |
470230-12 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted cholecystectomy surgical procedure, a large hem-o-lok clip applier instrument would not clamp clip.The procedure was completed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the large hem-o-lok clip applier instrument to perform failure analysis (fa).Fa was able to confirm/reproduce the reported complaint.The instrument was found to have grip input disk broken.Input disk 6 was found completely detached from the base of the housing.The instrument was found to have hairline cracks on grip input disk #7.The probable root cause is attributed to use and reprocessing conditions.The input disk can be susceptible to chemical or thermal stresses, which can result in the appearance of cracks in the ultem or peek material.Under repeated or prolonged chemical or thermal exposure cycles, these cracks can propagate into larger cracks, and may cause the input disk to break and separate from the instrument.
|
| |
|