| Model Number |
0672 |
| Medical Device Problem Codes |
Device Sensing Problem (2917); Impedance Problem (2950); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited sensing and impedance anomalies during testing performed after implantation.Subsequently, multiple attempts were made by the physician to reposition the lead; however, acceptable electrical measurements could not be obtained.The procedure was ultimately completed using a different lead, after which stable sensing and impedance measurements were achieved.This product has not been returned for analysis.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|