• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Xomed Inc. TRICUT® STRAIGHT SHAFT 4MM 11CM LONG; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Xomed Inc. TRICUT® STRAIGHT SHAFT 4MM 11CM LONG; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1884004
Medical Device Problem Code Vibration (1674)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to use, the blade was not rotating, and the inner blade was bent.There was no patient involvement.
 
Additional Manufacturer Narrative
H3: the tubing set, device, and an open pouch were returned in a large plastic sleeve.Upon receipt, the pouch was open on two of its four sides.Visual examination revealed no contamination on the tubing set or the blade.The inner assembly was found separated from the outer assembly upon return.No damage was noted to the tubing set, outer tube, and the inner and outer hubs.However, the inner shaft was found to be bent.The inner assembly was inserted into the outer assembly and rotated freely by hand without binding.The device was then loaded into a handpiece and, while oscillating up to 5,000 rpm, exhibited wobbling.The device failed due to the defective condition upon return and the inability to rotate properly.In the returned condition, the inner shaft was found to be bent, and the device was wobbling; therefore, the complaint was confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRICUT® STRAIGHT SHAFT 4MM 11CM LONG
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
Medtronic Xomed Inc.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
Medtronic Xomed Inc.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
+1(763)526-5677
MDR Report Key50911082
Report Number1045254-2026-01314
Device Sequence Number12213233
Product Code EQJ
UDI-Device Identifier00763000543846
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1884004
Device Catalogue Number1884004
Device Lot Number0231318664
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/25/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-