On (b)(6) 2026, a patient underwent an angioplasty procedure using a conquest 40 pta balloon.During the procedure, the balloon was successfully inflated to dilate a stent.After deflation and withdrawal for repositioning to another treatment site, the physician attempted to reinflate the balloon but observed that it would not maintain pressure.There was no reported patient injury.
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample has not been returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) are adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow up report will be submitted as applicable.D2b (lit; dqy)
section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|