| Model Number |
0998-00-0800-55 |
| Medical Device Problem Codes |
Battery Problem (2885); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) had fault a4.There was no patient involvement, no harm reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that the error is not reproducible.The system simply requested battery maintenance tests via a message.Once the tests were performed, the message stopped appearing.Maintenance was performed on batteries.Product passed all functional and safety tests per manufacturer specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site name:(b)(6) hospital.Contact person name:(b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|