H3: product analysis: evaluation of the device confirmed device seized.The likely cause of failure was detached distal bearings.It was also noted that pre temperature test could not be performed, laser markings were faded and color ring was faded.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|