CAREFUSION 303, INC. BD Pyxis¿ MedBank Mini; Automated Dispensing Cabinet Back to Search Results
Model Number 500004500500
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Application Program Problem (2880)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using the bd pyxis¿ medbank mini the station was stuck on blue screen.The customer attempted to reboot the machine and unplugged and replugged the power cable; however, the issue persisted.However, there were no delay or adverse events or injuries reported based on this event.
 
Additional Manufacturer Narrative
A review of the complaint history for sn (b)(6)was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6)was performed from the date of manufacture, 03-oct-2024 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the station was stuck on blue screen.The field service engineer (fse) found the station displaying a blue screen with a windows startup error and determined that the system could not boot correctly.The fse reseated the hard drive and attempted multiple recovery procedures, including safe mode, last known good configuration, and automated repair; however, the station remained unbootable.The fse then attempted to obtain a replacement hard drive, but the required part was not immediately available, resulting in the repair being postponed pending part acquisition.After a replacement hard drive became available, the fse installed the new hard drive, restored the station to operational status, and verified that the application launched successfully.The fse then transferred the station to the technical support center (tsc) for database rebuild activities.The issue was resolved, and the station was returned to service.The system functioned as intended after the field service engineer repaired the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Pyxis¿ MedBank Mini
Common Device Name
Automated Dispensing Cabinet
Manufacturer (Section D)
CAREFUSION 303, INC.
3750 torrey view ct
san diego CA 92130
Manufacturer (Section G)
CareFusion 303, INC.
10020 pacific mesa blvd
san diego CA 92121 -438
Manufacturer Contact
brett wilko
3750 torrey view ct
san diego, CA 92130
+1(858)617-2000
MDR Report Key50911626
Report Number2016493-2026-66577
Device Sequence Number17491217
Product Code BRY
UDI-Device Identifier10885403517181
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number500004500500
Device Catalogue Number1147-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/03/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-