| Model Number |
500004500001 |
| Medical Device Problem Codes |
Circuit Failure (1089); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medbank tower aux cabinet was rebooting unexpectedly at least five times per day.Followed these unexpected reboots, the cabinet did not consistently return to an online state automatically.In several instances, the customer was required to perform an additional manual reboot to restore functionality and bring the cabinet back online.However, there were no delays or adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
E.1.Initial reporter facility: (b)(6) hospital.A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 11-jul-2022 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that station displayed a black screen due to an intermittent connection issue.The field service engineer (fse) disconnected and reconnected the scanner usb cable.Coordinated with medbank support for remote diagnostics, which identified drawers 16 and 17 as non-functional.Shut down the device and replaced the controller cards for drawers 16 and 17, then rebooted the station.The system functioned as intended after the field service engineer repaired the device.
|
| |
|
Search Alerts/Recalls
|