• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. Bard, Inc. (BASD) -3006260740 Apheresis Port Access System; Port & Catheter, Implanted, Subcutaneous, Intravascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. Bard, Inc. (BASD) -3006260740 Apheresis Port Access System; Port & Catheter, Implanted, Subcutaneous, Intravascular Back to Search Results
Catalog Number A710962
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Obstruction of Flow (2423)
Health Effect - Clinical Codes Unspecified Infection (1930); Obstruction/Occlusion (2422); Thrombosis/Thrombus (4440)
Date of Event 09/14/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported through the litigation process that, sometime after the port placement, the patient was diagnosed with an unspecified infection, thrombosis and occlusion.However, the current status of the patient was unknown.
 
Additional Manufacturer Narrative
H11: b3.The date received by manufacturer has been used for this field.Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this lot.Investigation summary: the physical device was not returned for evaluation.No medical records were provided for review.The investigation is inconclusive for the reported infection, thrombosis and occlusion as no objective evidence was provided for review.A review of sterilization records, endotoxin testing results, and bioburden data was conducted in accordance with applicable quality system and regulatory requirements, and no nonconformances or anomalies were identified.Based upon available information, the definitive root cause was unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Apheresis Port Access System
Common Device Name
Port & Catheter, Implanted, Subcutaneous, Intravascular
Manufacturer (Section D)
C.R. Bard, Inc. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
Bard Reynosa S.A. de C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key50911647
Report Number3006260740-2026-07038
Device Sequence Number5752999
Product Code PTD
UDI-Device Identifier00801741129438
UDI-Public(01)00801741129438(17)260630(10)REKN3031
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K163001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Catalogue NumberA710962
Device Lot NumberREKN3031
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-