The physical device was not returned for investigation.The case description states two uncommand boluses of 10 u and 15 u resulted in hypoglycemia.The physical controller was not received for investigation.Ios log files from the complaint device were uploaded to the cloud system by the user and downloaded for investigation.Investigation of the ios log files confirmed the 10 u bolus was delivered on (b)(6) 2026 and 15 u bolus was delivered on (b)(6) 2026.The log files showed each bolus amount was typed in one digit at a time and confirm and start boluses were tapped.Each bolus had records of controller interactions in the log files.No evidence of an uncommand bolus was observed.Locked down smartphone: phone_control_ios,
omni pod software app version: 2.2.1,
operating system: 26.5,
hardware: iphone14.7,
cgm sensor type: g7.Please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
|