• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Datascope Corp. TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES; System, Balloon, Intra-Aortic And Control

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Datascope Corp. TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0684-00-0608
Medical Device Problem Code Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that after placement of the transray plus 40cc intra-aortic balloon (iab) catheter, pumping was started.Although it was confirmed on the display that pumping was being performed, the augmentation pressure did not increase at all.Repositioning of the catheter was attempted, but no improvement was observed.Therefore, the device was replaced with a izumi kogyo iab, after which improvement was noted.No harm to the patient was reported.
 
Additional Manufacturer Narrative
E.Initial reporter: event site postal code: (b)(6).Event site telephone: (b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50911787
Report Number2248146-2026-0001572
Device Sequence Number5370086
Product Code DSP
UDI-Device Identifier10607567111728
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/26/2025
Device Model Number0684-00-0608
Device Catalogue Number0684-00-0608
Device Lot Number3000485988
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/21/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Cardiosave IABP (b)(4).
Patient SexUnknown
-
-