• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Siemens Healthcare Diagnostics Inc. Atellica CH; Atellica CH Carbon Dioxide, Concentrated (CO2_c)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Siemens Healthcare Diagnostics Inc. Atellica CH; Atellica CH Carbon Dioxide, Concentrated (CO2_c) Back to Search Results
Model Number N/A
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that erroneously elevated carbon dioxide (co2, lot#160071) results were obtained on multiple patient samples on an atellica ch 930 analyzer.The initial results were reported to the physician(s) but not questioned.The samples were repeated on the original atellica ch 930 analyzer and obtained lower results which were considered correct and reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the erroneously elevated co2 results.
 
Additional Manufacturer Narrative
A united states (us) customer contacted a siemens customer care center (ccc) and reported that erroneously elevated carbon dioxide (co2_c) results were obtained on multiple patient samples on an atellica ch 930 analyzer.Siemens reviewed the available instrument data and determined that the co2_c calibration and quality control (qc) recovery patterns were consistent with a water quality issue beginning on 02-sep-2026 and resolving around 20-sep-2026.Quality control (qc) recovery was out of range on the day of the event.The customer performed a co2_c pack calibration (calibration id (b)(4)) following elevated qc recovery, after which qc recovered within the expected range.Siemens' review of calibration, qc, and patient data did not identify evidence of an analyzer or reagent issue.Some samples considered erroneous by the customer were not repeated until over 24 hours after the initial elevated results were obtained.Siemens determined that the lower repeat co2_c results may have been related to prolonged sample exposure to air and/or underfilled sample tubes prior to repeat testing.The specific cause of the water quality issue and the reported event could not be determined but most likely due to prolonged exposure to air prior to repeat testing being performed.The issue was resolved through routine troubleshooting and correction of the water quality condition.The customer is operational.No product issue was identified.The device is performing within specifications.No further evaluation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Atellica CH
Common Device Name
Atellica CH Carbon Dioxide, Concentrated (CO2_c)
Manufacturer (Section D)
Siemens Healthcare Diagnostics Inc.
511 benedict ave
tarrytown NY 10591
Manufacturer (Section G)
Siemens Healthcare Diagnostics Inc.
500 gbc drive, po box 6101
registration # 2517506
newark DE 19714 -610
Manufacturer Contact
varsha joshi
pritech park sez, block 11
sarjapur outer ring road
karnataka 56010-3
IN   560103
+919(870)577433
MDR Report Key50911831
Report Number2432235-2026-00230
Device Sequence Number5330823
Product Code KHS
UDI-Device Identifier00630414595900
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K100289
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number11097521
Device Lot Number160071
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/02/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-