B3: date of incident or estimated date of incident was not provided to the manufacturer.Event notified date used as date of incident until further information is obtained.G3: aware date used as date of analysis performed date as the event is reported based on evaluation findings.H3: product analysis: evaluation of the device determined that the reported malfunction of gripped rust could not be confirmed.It was also noted that the tube hole was oval, the distal bearing was detached, the pre-repair temperature test could not be performed, the laser markings were partially illegible, and the painted color ring was faded.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|