A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 22-jun-2026 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the drawers 04 and 05 being incorrectly configured in touch mode, which prevented the drawers from opening when users attempted to retrieve medications.A technical support specialist (tss) changed the drawer configuration from touch mode to normal zone mode, and the user was instructed to log back in.After the configuration change, the user was able to access medications from drawer 05 successfully, confirmed that the issue was resolved.The system functioned as intended after the technical support specialist troubleshot the device.
|