| Catalog Number |
25-40-45 |
| Medical Device Problem Code |
Gas Output Problem (1266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Livanova deutschland received a report of an electronic gas blender (egb) for which an error message was displayed and could not operate properly.The event occurred during priming.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A.1.-a.5.Patient information was not provided.H11:
a loan unit was provided in substitution to involved device.Additionally, the customer reported that the issue occurred a few days after device was returned from a different service activity.A livanova field service engineer (fse) was dispatched on site and the event was not confirmed.The gas blender involved was tested for one day, with different parameters being varied during the evaluation, and no error nor anomaly was observed.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|