• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT Puerto Rico Operations Co, Med Rel CD HORIZON® Fenestrated Screw Spinal System; BONE CEMENT, POSTERIOR SCREW AUGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT Puerto Rico Operations Co, Med Rel CD HORIZON® Fenestrated Screw Spinal System; BONE CEMENT, POSTERIOR SCREW AUGMENTATION Back to Search Results
Model Number 55750026545
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from the health care professional via a manufacturer representative regarding a device used for l2-5 lateral lumbar interbody fusion, l2-l5 ercutaneous pedicle screw fixation, l2, l4, l5 femoral neck system procedures.The preoperative diagnosis for this surgery was l2/3/4/5 lumbar spinal canal stenosis with degenerative scoliosis.It was reported that during intra-op,when filling the cement, it felt harder than usual.During cement insertion on the right side of l2, there was a hard feeling, but imaging was also confirmed and filling was continued as is.Cement leakage was confirmed when the device was removed.Therefore, the right l2 screw was replaced with a standard mas.After removal, it was confirmed that the cement had not reached the hole at the tip of the screw and that almost all of it had leaked into the head.The device had been inserted slightly obliquely, and the confirmation button was also in a state where it could be pressed slightly.There were no patient symptoms reported.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® Fenestrated Screw Spinal System
Common Device Name
BONE CEMENT, POSTERIOR SCREW AUGMENTATION
Manufacturer (Section D)
MDT Puerto Rico Operations Co, Med Rel
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer (Section G)
MDT Puerto Rico Operations Co, Med Rel
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
+1(763)526-5677
MDR Report Key50912801
Report Number2647346-2026-00547
Device Sequence Number5776403
Product Code PML
UDI-Device Identifier00763000148836
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K201362
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number55750026545
Device Catalogue Number55750026545
Device Lot NumberH6046203
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/28/2026
Date Device Manufactured04/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-