| Model Number |
SC-1232 |
| Medical Device Problem Code |
Unexpected Therapeutic Results (1631)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
03/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient was experiencing inadequate stimulation.The patient underwent a procedure wherein the ipg and spinal cord stimulator (scs) leads were explanted.The patient was doing well postoperatively and the explanted devices were retained by the medical facility and will not be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional suspect medical device components involved in the event:
model number/catalog number: sc-2218-50
serial number: (b)(6)
batch/lot number: 7104214
model/catalog description: linear st lead kit 50 cm
unique identifier (udi)#: (b)(4).Model number/catalog number: sc-2218-50
serial number: (b)(6)
batch/lot number: 7104321
model/catalog description: linear st lead kit 50 cm
unique identifier (udi)#: (b)(4).Model number/catalog number: sc-4318
batch/lot number: 28820370
model/catalog description: linear st lead kit 50 cm
unique identifier (udi)#: (b)(4).Investigation results:
all explanted devices were not returned for analysis; therefore, a technical analysis could not be performed.Device history record:
it was confirmed that all devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review:
review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion:
all explanted devices were not returned for analysis and the complaint as reported could not be confirmed.Record review revealed no additional information related to all complaints.Therefore, this investigation is unable to determine a probable cause for the complaint and the conclusion code cause not established will be used.
|
| |
|