• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION WaveWriter Alpha?; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION CORPORATION WaveWriter Alpha?; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-1232
Medical Device Problem Code Unexpected Therapeutic Results (1631)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was experiencing inadequate stimulation.The patient underwent a procedure wherein the ipg and spinal cord stimulator (scs) leads were explanted.The patient was doing well postoperatively and the explanted devices were retained by the medical facility and will not be returned.
 
Additional Manufacturer Narrative
Additional suspect medical device components involved in the event: model number/catalog number: sc-2218-50 serial number: (b)(6) batch/lot number: 7104214 model/catalog description: linear st lead kit 50 cm unique identifier (udi)#: (b)(4).Model number/catalog number: sc-2218-50 serial number: (b)(6) batch/lot number: 7104321 model/catalog description: linear st lead kit 50 cm unique identifier (udi)#: (b)(4).Model number/catalog number: sc-4318 batch/lot number: 28820370 model/catalog description: linear st lead kit 50 cm unique identifier (udi)#: (b)(4).Investigation results: all explanted devices were not returned for analysis; therefore, a technical analysis could not be performed.Device history record: it was confirmed that all devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: all explanted devices were not returned for analysis and the complaint as reported could not be confirmed.Record review revealed no additional information related to all complaints.Therefore, this investigation is unable to determine a probable cause for the complaint and the conclusion code cause not established will be used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WaveWriter Alpha?
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
ricardo amaro castro
25155 rye canyon loop
valencia, CA 91355
+1(320)800-7570
MDR Report Key50912854
Report Number3006630150-2026-07139
Device Sequence Number5778539
Product Code LGW
UDI-Device Identifier08714729985099
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date08/26/2023
Device Model NumberSC-1232
Device Catalogue NumberSC-1232
Device Lot Number521394
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/28/2026
Date Device Manufactured08/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient SexFemale
-
-