| Model Number |
FX2200X15B |
| Medical Device Problem Codes |
Circuit Failure (1089); Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received a report from a customer that a ventilator inoperative alarm occurred on a trilogy ev300 ventilator.No patient harm or injury was reported.The ventilator was evaluated by a field service engineer (fse).The system board was suspected to be defective.Once the system board was replaced and the software was reloaded, the ventilator was operating properly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of vent inop is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Device logs were not available; therefore, the date of the most recent vent inop occurrence could not be determined.Accordingly, the date the device was evaluated has been used as the event date for reporting purposes.Dhr review was performed for the complaint device sn.Review confirms that the device met the final acceptance criteria and was released for final distribution.
|
| |
|
Search Alerts/Recalls
|