The account reported finding error message, and it was reported that the thermometers did not function as expected and that an error code occurred each time the devices were used.A photograph and product samples were provided; review of the photograph showed the device was observed to power on, after which error 1 appeared on the display, and two returned samples displayed error 1 during startup before control batteries were installed and valid readings were obtained.Complaint was confirmed.Finished good production records reviewed found no abnormalities, the cause was documented as cannot determine cause with received sample, and after replacing the batteries and allowing the devices to remain powered off for a short duration, the thermometers functioned as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|