The customer reported that the mattress does not hold air and that the top mattress baldders is not inflating.The sample provided was only the pump, which was attached to a mattress used for testing; the mattress was fully inflated and the green normal pressure light stayed on even with added 175lbs of weight.Medline was unable to find any quality defects with the pump, and the cause was not confirmed with the received sample.No root cause was identified; there was no reported death or serious injury associated with this event, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|