• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION 303, INC. BD Pyxis¿ MedStation¿ ES Tower; Automated Dispensing Cabinet

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION 303, INC. BD Pyxis¿ MedStation¿ ES Tower; Automated Dispensing Cabinet Back to Search Results
Model Number 500001002506
Medical Device Problem Code Application Program Freezes or Fails to Launch (4031)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using the bd pyxis¿ medstation¿ es tower, the system had a medical device correction issue.The customer stated that this malfunction caused a delay in dispensing medication to patients.However, there were no adverse events or injuries reported based on this event.
 
Additional Manufacturer Narrative
A review of the complaint history for sn: (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn: (b)(6) was performed from the date of manufacture, 24-jun-2025 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the station had medical device correction issue.A technical support specialist (tss) addressed issues mms-25-5334 and mms-25-5335 by following knowledge article "fa mms 25-5334 & 25-5335 delayed and missing patient and/or order information & observer tool".Transferred the required files and installed the bd inbound observer tool through task scheduler to monitor and remediate the inbound message processing issue.Confirmed the installation completed successfully and notified the customer through email that the remediation had been completed.The issue was resolved, and the case was closed.The system functioned as intended after the technical support specialist troubleshot the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Pyxis¿ MedStation¿ ES Tower
Common Device Name
Automated Dispensing Cabinet
Manufacturer (Section D)
CAREFUSION 303, INC.
3750 torrey view ct
san diego CA 92130
Manufacturer (Section G)
Sistemas Medicos Alaris S.A de C.V.
parque sur, el realito
blvd. 2000
tijuana 22250
MX   22250
Manufacturer Contact
brett wilko
3750 torrey view ct
san diego, CA 92130
+1(858)617-2000
MDR Report Key50913267
Report Number2016493-2026-66692
Device Sequence Number14897195
Product Code BRY
UDI-Device Identifier10885403512674
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number500001002506
Device Catalogue Number352
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/24/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-