A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 06-may-2026 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the user was unable to issue the item.A technical support specialist (tss) reviewed the medication transaction and identified an incorrect medication count.The tss worked with nursing staff, provided guidance on performing a correct cycle count, and corrected the medication count.The tss then assisted with overriding and issuing the medication, coordinated with pharmacy for verification, and confirmed that the medication was successfully issued.The system functioned as intended after the technical support specialist troubleshot the device.
|