A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 23-jan-2020 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that medbank application became unresponsive.The technical support specialist (tss) closed the medbank application, verified universal serial bus (usb) power management settings were already turned off in device manager, restarted the medbank application, and logged back into the system.The system functioned as intended after the technical support specialist (tss) troubleshot the issue.
|