Due to the limited information and the missing product investigation is restricted to traceability of manufacturing and quality control data.Based on the production records, a defect at the time of delivery of the mininav is excluded.The traceability indicates that the shunt was in perfect condition.Regarding similar complaints where we could examine the samples, it could be determine that the most frequent cause for a deterioration in the function of the product is deposits caused by natural substances in the cerebrospinal fluid,
e.G.Protein, blood or tissue particles.These are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve if aggregated in the valve.A final clarification of the cause what has led to drainage problems is only possible by an examination.
|