A complaint was received related to a "tearing when donning" associated with the sensicare pi surgical glove, and the complaint transcript reported "ripping when scrubber audra was donning gloves." a sample and image were provided with the torn cuff, trending was performed over the last 2 years of complaints, the lot inspection was completed with passing results, and the vendor's review of batch records and retention samples found no abnormalities or deviations.The complaint was able to be confirmed via the returned sample, but no root cause could be determined.No adverse trending was identified at this time, and the review indicated no need for further action.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|