| Model Number |
0998-00-0800-33 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a routine check the cardiosave intra-aortic balloon pump (iabp) is showing the error message \"internal communication failure\".No patient was involved.No harm was reported.A getinge field service engineer was dispatched to evaluate the unit.The fse reset the connections for supply board in the cart, connections for power management, solenoid driver and all other electronics.The iabp was ran on test balloon for 24 hours, no issue was observed.Inspection is complete.The unit passed all functional and safety tests to manufacturer specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
E1.Event site name: (b)(6).\
e1.Event site postal code: (b)(6).\
\
upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|