The serial number of the customer's e 801 analyzer is (b)(6).\
\
the serial number of the e 801 analyzer used for investigation is (b)(6).\
\
single false reactive results can occur based on the labeled specificity claim of the assay.\
\
for repeatedly reactive samples, cdc recommendations for supplemental testing must be followed.\
\
for diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings.\
\
the investigation is ongoing.
|
|
The initial reporter stated they received discrepant results for one patient sample tested with elecsys anti-hcv ii on\t\
two cobas e 801 analytical units.\
\
the sample was tested twice on the customer's e 801 analyzer, resulting in anti-hcv values of 26.4 coi (reactive) and 26.5 coi (reactive).\
\
the sample was provided for investigation, where it was tested twice on a second e 801 analyzer on 10-sep-2026, resulting in anti-hcv values of 26.0 coi (reactive) and 26.1 coi (reactive).\
\
during investigations, the sample was tested with an unknown clia assay method, resulting in an anti-hcv value of 0.67 (negative).No units of measure were provided.\
\
during investigations, the sample was also tested with the fujirebio hcv score method, resulting in the following values:\
c1 antigen = negative.\
c2 antigen = negative.\
e2 antigen = negative.\
ns3 antigen = \"+/-\".\
ns4 antigen = negative.\
ns5 antigen = negative.\
final result = inconclusive.
|