• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Ireland Micra¿ AV2; LEADLESS PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Ireland Micra¿ AV2; LEADLESS PACEMAKER Back to Search Results
Model Number MC2AVR1
Medical Device Problem Codes Failure to Capture (1081); Under-Sensing (1661); Capturing Problem (2891); Positioning Problem (3009); Unstable Capture Threshold (3269)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the implant procedure of a leadless implantable pulse generator (ipg) the leadless ipg exhibited non-cap ture, undersensing and out of range (oor) thresholds.Recapturing difficulty was also noted.The leadless ipg was successfully recaptured and attempted/ not used.No patient complications have been reported as a result of this event.
 
Event or Problem Description
It was further reported that the leadless ipg exhibited unstable thresholds.Once the tether was cut for the first deployment the th resholds were fine, but subsequently rose and kept rising.The ipg was removed with a snare.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Continuation of d10:this is a system report.The section d information is for the primary device, which was in use with the following: brand name "[micra]" product id "[mc2avr1-delsys]" (serial: (b)(6).D9: yes, return date: 29 sep 2026 h3: <(><<)>no> medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Micra¿ AV2
Common Device Name
LEADLESS PACEMAKER
Manufacturer (Section D)
Medtronic Ireland
parkmore business park west
galway
EI 
Manufacturer (Section G)
Medtronic Ireland
parkmore business park west
galway
EI  
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50913678
Report Number9612164-2026-04858
Device Sequence Number17491232
Product Code PNJ
UDI-Device Identifier00763000078645
UDI-Public(01)00763000078645(17)271028
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P150033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2AVR1
Device Catalogue NumberMC2AVR1
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/29/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2026
Supplement Date Received by Manufacturer09/28/2026
Initial Report FDA Received Date09/28/2026
Supplement Report FDA Received Date10/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/08/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexFemale
-
-