| Model Number |
500001002506 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medstation¿ es tower, biometric identification (bio-id) was failed.However, there were no delay to patient care and adverse events or injuries reported based on this incident.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn: (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn: (b)(6) was performed from the date of manufacture, 07-nov-2023 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the bio-ids were not installed per sop, as the devices were connected before the drivers were properly installed.A field service engineer (fse) uninstalled the existing drivers and reinstalled the updated drivers following the correct sop.After completing the repair, the customer performed multiple login attempts using bio-id with no further failures observed.The system functioned as intended after the field service engineer repaired the device.
|
| |
|