| Model Number |
GIF-1100 |
| Medical Device Problem Codes |
Erratic or Intermittent Display (1182); No Display/Image (1183); Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported black image and image intermittent during the reprocessing involving an olympus gastrointestinal videoscope.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was not confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: objective lens dirty and blurry image.A root cause could not be identified.Based on the results of the investigation.The most probable cause of the complaint was cause not determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|