• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Respironics, Inc. Trilogy 100; Ventilator, Continuous, Facility Use

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Respironics, Inc. Trilogy 100; Ventilator, Continuous, Facility Use Back to Search Results
Model Number LA1054096
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received a report that a trilogy 100 ventilator was returned for evaluation.No patient injury or serious harm was reported.The device was evaluated at a manufacturer service center.A visual inspection was performed, and no evidence of foam particles or foam degradation was observed.The device failed to power on during assessment.In addition, the data was unable to be read as there was no access to the information.No documentation regarding component replacement was available.No repair activities were performed.The device was scrapped.A final report is being submitted.If any additional information is received, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The reported failure of the device would not power on is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review not required.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Trilogy 100
Common Device Name
Ventilator, Continuous, Facility Use
Manufacturer (Section D)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
+1(724)733-0200
MDR Report Key50913881
Report Number2518422-2026-046503
Device Sequence Number5344778
Product Code CBK
UDI-Device Identifier00606959022836
UDI-Public(01)00606959022836
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberLA1054096
Device Catalogue NumberLA1054096
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/04/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-