A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 25-may-2022 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the device was in disconnect mode and on critical override.A field service engineer (fse) worked with site it to troubleshoot recurring disconnected mode and critical override events following scheduled reboots.Verified network infrastructure, reseated internal ethernet connections, replaced the all in one and lan cable, corrected network speed and duplex settings, and confirmed cabling integrity with it.Despite these actions, the station continued to enter disconnected mode after reboot.A tsc lead then connected to the station, restarted the data synchronization service, and rebooted the system.Following the reboot, communication was restored, synchronization resumed normally, and the station returned to normal operation.The system functioned as intended after the field service engineer troubleshot the device.
|