The customer reported several gloves that were taken out of the packages had holes in them.Inspection reports were reviewed, the lot passed inspection for all physical properties per astm standards, and a historical complaint query for the last 24 months identified no trends; no sample was provided, and one photo of the reported issue was received.The issue is confirmed, and the manufacturer identified the probable cause to be thin spots.The manufacturer will conduct refreshment visual training, and based on the investigation results, no further actions are being taken at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|