The account reported finding defective bovie, and the end user reported that the cut function would activate even when the button is not pushed; the issue occurred during an abdominal procedure with patient involvement and no injuries.Medline received one used cautery pencil; the rest of the pack, cautery holster, and blade tip were not received, and there was no visual damage or defects with the cautery pencil, cord, or plug end.The vendor response stated that visual inspection found no notable conditions, functional testing found the pencils did not self-activate and functioned normally when activated in the generator, and the reported condition was not confirmed.The primary analysis found no device problem, the product functioned as intended, and the most likely root cause could not be determined.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|