Model Number 1340
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the power module, ppm-21599, had a yellow wrench.The alarm was unable to be cleared.The battery was reseated and changed to a new backup battery with no resolution.The power module was planned to be returned for repair.
 
Additional Manufacturer Narrative
Section a: there was no patient involvement.Section d4: device lot number were not provided, and expiration date and manufacture date (h4) cannot be determined.The primary udi number in d4 is reflective of all available information.No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE® POWER MODULE (N/A)
Common Device Name
Ventricular (Assist) Bypass
Manufacturer (Section D)
Thoratec Corporation
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
Thoratec Corporation
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
fredric gold
6035 stoneridge drive
pleasanton, CA 94588
+1(781)852-8207
MDR Report Key50914280
Report Number2916596-2026-2919788
Device Sequence Number5344791
Product Code DSQ
UDI-Device Identifier00813024010654
UDI-Public(01)00813024010654
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P160054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1340
Device Catalogue Number1340
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-