MAUDE Adverse Event Report: Respironics, Inc. Respironics; Ventilator, Continuous, Minimal Ventilatory Support, Facility Use

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Respironics, Inc. Respironics; Ventilator, Continuous, Minimal Ventilatory Support, Facility Use Back to Search Results
Model Number V60 V60Plus Ventilator
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
On 16sep2026, philips received a complaint from the second provincial general hospital of guangdong province on the v60 indicating a device malfunction as the device issued a "proximal pressure sensor autozero failed" alarm while the department was preparing to use this ventilator.It was reported that there was no patient involvement at the time the issue was discovered.No patient or user harm was reported.Normal operation resumed after the device was replaced.Based on the case information provided, the device did not meet product specifications.The alleged malfunction impacted the user¿s ability to use the device properly.This investigation is ongoing.
 
Additional Manufacturer Narrative
E1: reporting institution phone #: (b)(6).Reporter phone #: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Respironics
Common Device Name
Ventilator, Continuous, Minimal Ventilatory Support, Facility Use
Manufacturer (Section D)
Respironics, Inc.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
+1(724)733-0200
MDR Report Key50914316
Report Number2518422-2026-115476
Device Sequence Number5544683
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public(01)00884838025776
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K102985
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberV60 V60Plus Ventilator
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2026
Initial Report FDA Received Date09/28/2026
Date Device Manufactured11/06/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-