• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZimVie US Corp LLC T3® PRO TAPERED IMPLANT 5/4MM (D) X 8.5MM (L); Dental Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZimVie US Corp LLC T3® PRO TAPERED IMPLANT 5/4MM (D) X 8.5MM (L); Dental Implant Back to Search Results
Catalog Number T3PT5485
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377)
Date of Event 08/12/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported peri-implantitis, when switching the cover screw for an encode healing abutment for #19,20 implant, fair amount of granulation was found around both implants and they were loose.Implant had to be removed.This event refers to the peri-implantitis in tooth site 19.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4) d9: device availability unknown / not provided d10.Concomitant medical products t3pt4385, t3® pro tapered implant 4/3mm (d) x 8.5mm (l) lot 2120029404 a summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T3® PRO TAPERED IMPLANT 5/4MM (D) X 8.5MM (L)
Common Device Name
Dental Implant
Manufacturer (Section D)
ZimVie US Corp LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZimVie US Corp LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
+1(561)971-3230
MDR Report Key50914383
Report Number0001038806-2026-04909
Device Sequence Number14788513
Product Code DZE
UDI-Device Identifier00844868046806
UDI-Public(01)00844868046806(17)300828(10)2120037177
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K213672
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberT3PT5485
Device Lot Number2120037177
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Dental Implant See H10 Narrative
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexFemale
Patient Weight165 LB
-
-