It was reported that during service and evaluation, it was determined that the meniscal deployment gun device spring pusher tip was deformed.It was noted in the service order that the device could not fix the meniscus after deployment during a meniscal repair surgical procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative:
the product was returned to depuy synthes for evaluation, and an investigation was conducted in accordance with depuy synthes procedures.Visual inspection of the returned device found the spring pusher tip was deformed.The spring pusher had foreign matter, presumable biological matter.No other anomalies were noted.A functional test was unable to be performed due to the post-manufacturing damage.The reported complaint was confirmed based on the evaluation results.The investigation traced the potential cause to aggressively inserting the needle on the gun's connector and rough maneuvering of the device.As per ifu, it is necessary to use a calibrated probe, measure the width of the meniscal tissue to be repaired and set the adjustable depth.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.There was no non-conformance regarding this lot.
|