• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DePuy Mitek LLC US MENISCAL DEPLOYMENT GUN; Applier, Staple, Surgical,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DePuy Mitek LLC US MENISCAL DEPLOYMENT GUN; Applier, Staple, Surgical, Back to Search Results
Catalog Number 228143
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during service and evaluation, it was determined that the meniscal deployment gun device spring pusher tip was deformed.It was noted in the service order that the device could not fix the meniscus after deployment during a meniscal repair surgical procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: the product was returned to depuy synthes for evaluation, and an investigation was conducted in accordance with depuy synthes procedures.Visual inspection of the returned device found the spring pusher tip was deformed.The spring pusher had foreign matter, presumable biological matter.No other anomalies were noted.A functional test was unable to be performed due to the post-manufacturing damage.The reported complaint was confirmed based on the evaluation results.The investigation traced the potential cause to aggressively inserting the needle on the gun's connector and rough maneuvering of the device.As per ifu, it is necessary to use a calibrated probe, measure the width of the meniscal tissue to be repaired and set the adjustable depth.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.There was no non-conformance regarding this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENISCAL DEPLOYMENT GUN
Common Device Name
Applier, Staple, Surgical,
Manufacturer (Section D)
DePuy Mitek LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
+1(908)386-3534
MDR Report Key50914647
Report Number1221934-2026-04373
Device Sequence Number5393531
Product Code GEF
UDI-Device Identifier10886705010059
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number228143
Device Lot Number10BMSU
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/15/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-