| Model Number |
WA50052A |
| Medical Device Problem Codes |
No Display/Image (1183); Dent in Material (2526); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the rigid video laparoscope image went out during an inspection for use.During the initial evaluation of the device, dents and scratches were observed on the distal end.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection, and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: dents and scratches on distal end.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: no image caused by defected outer tube.The most probable cause of dents and scratches on distal end is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|