| Model Number |
480422-03 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Unspecified Tissue Injury (4559)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted right hemicolectomy surgical procedure, while using the vessel sealer extend (vse) instrument, the instrument was not sealing and gave an error indicating a non-complete sealing cycle.Tissue was adhered to the instrument which caused tissue damage and bleeding.Tissue was never cut that was not fully sealed and the instrument was noted to work ¿sometimes¿ with audible tones.The procedure was converted to laparoscopic with no additional ports placed.Information surrounding how the patient tolerated the change in surgery was not known.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did not receive the vessel sealer instrument to perform failure analysis.A field service engineer (fse) also replaced the e-200 generator due to the issue.Vessel sealer logs show this vessel sealer extend (vse) was installed two times on universal surgical manipulator (usm) arm 2 and one time on usm arm 4.The instrument was used for approximately 42 minutes.E-200 activations logs were unavailable.
|
| |
|