It was reported that approximately three weeks post implant of an implantable pulse generator (ipg) system, the patient experienced asystole, which seen from the ambulance electrocardiogram (ecg) recording when they were hooked up to an external pacer.The patient also experienced bradycardia, dizziness and pre-syncope.The right ventricular (rv) lead exhibited unstable thresholds, no capture, oversensing, short v-v intervals, cross talk oversensing, undersensing, high and rising thresholds and high impedance.A possible micro perforation was suspected and the lead had pulled back resulting in poor tissue interface.The rv lead was explanted and replaced.It was also noted that the right atrial (ra) lead exhibited cross talk oversensing.The lead remains in use. no further patient complications have been reported as a result of this event.
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|