MAUDE Adverse Event Report: Medtronic Puerto Rico Villalba SelectSecure MRI SureScan; DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Puerto Rico Villalba SelectSecure MRI SureScan; DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) Back to Search Results
Model Number 383069
Medical Device Problem Codes Failure to Capture (1081); High impedance (1291); Over-Sensing (1438); Under-Sensing (1661); Capturing Problem (2891); Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993); High Capture Threshold (3266); Unstable Capture Threshold (3269)
Health Effect - Clinical Codes Bradycardia (1751); Dizziness (2194); Cardiac Perforation (2513); Asystole (4442)
Date of Event 08/25/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately three weeks post implant of an implantable pulse generator (ipg) system, the patient experienced asystole, which seen from the ambulance electrocardiogram (ecg) recording when they were hooked up to an external pacer.The patient also experienced bradycardia, dizziness and pre-syncope.The right ventricular (rv) lead exhibited unstable thresholds, no capture, oversensing, short v-v intervals, cross talk oversensing, undersensing, high and rising thresholds and high impedance.A possible micro perforation was suspected and the lead had pulled back resulting in poor tissue interface.The rv lead was explanted and replaced.It was also noted that the right atrial (ra) lead exhibited cross talk oversensing.The lead remains in use. no further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SelectSecure MRI SureScan
Common Device Name
DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV)
Manufacturer (Section D)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50914834
Report Number2649622-2026-28588
Device Sequence Number5999032
Product Code NVN
UDI-Device Identifier00763000740269
UDI-Public(01)00763000740269(17)280625(20)01
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P030036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number383069
Device Catalogue Number383069
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/07/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
407652 LEAD, W3DR01 IPG.
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age74 YR
Patient SexMale
Patient Weight224 LB
-
-