|
Synthes GmbH UNIUM¿ MODULAR HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
05.001.601 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the modular handpiece device had missing components.It was noted in the service order that the device was missing a ring.This event did not occur during surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional narrative:
investigation summary: the actual device was returned for evaluation.A visual and functional assessment was performed on the device.The described failure by the customer was confirmed.The device was visually inspected as received from the customer.It was found that the device failed visual inspection due to fell of heat cover.Following test failed according to the pre-diagnostic assessment:
general condition: fail.During the assessment it was found that the proactive heat cover fell off the handpiece and was missing.This is related to the missing recess in the housing for the heat protective cover.The most probable cause for the found defect is related to a manufacturing related issue by the supplier.Further investigation and assessment are addressed within our quality management system.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|