Model Number 1002717
Medical Device Problem Code Pumping Stopped (1503)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that pump displayed malfunction alarm for speaker and issue did not follow any notable event.There was no adverse impact to the customer¿s blood glucose level.Customer switched to multiple daily injections for insulin, and technical support arranged replacement pump with updated software, confirmed shipping details, instructed customer to place malfunctioning pump in storage mode and return it within 10 days, and advised customer to reprogram pump settings and reconnect continuous glucose monitor to replacement pump.
 
Additional Manufacturer Narrative
In use at the time of the event: cgm model: unknown.Mobile os: android / android_pixel 8 pro / unknown / android 16.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
t:slim X2 Insulin Pump with Control-IQ Technology
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Tandem Diabetes Care
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
john connell
12400 high bluff drive
san diego, CA 92130
+1(858)224-6281
MDR Report Key50915093
Report Number3013756811-2026-175766
Device Sequence Number17485300
Product Code QFG
UDI-Device Identifier00389152286716
UDI-Public(01)00389152286716(22)1012962(11)231212
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K203234
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1002717
Device Catalogue Number1012867
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/12/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age40 YR
Patient SexMale
-
-